Accuracy
A comparison was conducted using the KET One Step Ketamine Test and GC/MS at 1,000 ng/mL cut-off. The following results were tabulated: Method | GC/MS | Total Results | ||
KET One Step Test | Results | Positive | Negative | |
Positive | 10 | 8 | 18 | |
Negative | 0 | 205 | 205 | |
Total Results | 10 | 213 | 223 | |
% Agreement | 100% | 96.2% | 96.4% |
Analytical Sensitivity
A drug-free urine pool was spiked with Ketamine at the following concentrations: 0 ng/mL, 500ng/mL, 750 ng/mL, 1,000 ng/mL, 1,250 ng/mL and 1,500 ng/mL.The result demonstrates >99% accuracy at 50% above and 50% below the cut-off concentration. The data are summarized below: Ketamine Concentration (ng/mL) | Percent of Cut-off | n | Visual Result | |
Negative | Positive | |||
0 | 0 | 20 | 20 | 0 |
500 | -50% | 20 | 20 | 0 |
750 | -25% | 20 | 15 | 5 |
1,000 | Cut-off | 20 | 10 | 10 |
1,250 | +25% | 20 | 0 | 20 |
1,500 | +50% | 20 | 0 | 20 |
Analytical Specificity
The following table lists compounds that are positively detected in urine by the KET One Step Ketamine Test at 5 minutes.Compound | Concentration (ng/mL) |
Ketamine | 1,000 |
Pentobarbital | 50,000 |
Secobarbital | 100,000 |
Norketamine | 50,000 |